COVID-19 DRUGS FAVIPIRAVIR HPLC -UV METHOD VALIDATION PROCESS
Favipiravir
6-FLUORO-3-HYDROXYPYRAZINE-2-CARBOXAMIDE
IUPAC NAME- 5-fluoro-2-oxo-1H-pyrazine-3-carboxamide
Molecular Weight - 157.1 g/mol Favipiravir |
DISCRIPTION -
Favipiravir (FVP), a pyrazine analog, has shown antiviral activity against a wide variety of viruses. It is considered to be worth further investigation as a potential candidate drug for COVID-19. It is not officially available in any pharmacopoeia. A rapid, simple, precise, accurate, and isocratic high performance liquid chromatography (HPLC) method has been developed for routine quality control of favipiravir in pharmaceutical formulations. Separation was carried out by C18 column. The mobile phase was a mixture of 50 mM potassium dihydrogen phosphate (pH 2.3) and acetonitrile (90:10, v/v) at a flow rate of 1 mL min−1. The ultraviolet (UV) detection and column temperature were 323 nm, and 30 °C, respectively. The run time was 15 min under these chromatographic conditions. Excellent linear relationship between peak area and favipiravir concentration in the range of 10–100 μg mL−1 has been observed (r2, 0.9999). Developed method has been found to be sensitive (limits of detection and quantification were 1.20 μg mL−1and 3.60 μg mL−1, respectively), precise (the interday and intraday relative standard deviation (RSD) values for peak area and retention time were less than 0.4 and 0.2%, respectively), accurate (recovery, 99.19–100.17%), specific and robust (% RSD were less than 1.00, for system suitability parameters). Proposed method has been successfully applied for quantification of favipiravir in pharmaceutical formulations.
Introduction
Chinese-borne coronavirus disease (COVID-19) spread rapidly and became an epidemic, affecting almost all countries and regions around the world. COVID-19 case death rate ranges from 1% to 7% according to the reports of World Health Organization (WHO). It caused all people in the world to change their lifestyle. It still threatens the entire World [1]. Since the outbreak of the COVID-19 began to affect the world, countries have implemented different treatment methods.
Active therapeutic alternatives are urgently needed as a rising COVID-19 pandemic and possible effects on global health [2]. Many medications such as chloroquine, arbidol, remdesivir, and favipiravir are currently undergoing clinical trials in several countries to assess their effectiveness and safety in treating coronavirus disease [3, 4]. So far, there is no gold standard for the treatment of COVID-19 since there is not enough evidence [5].
Favipiravir (6-fluoro-3-hydroxypyrazine-2-carboxamide) is an analog of pyrazine (Fig. 1). Favipiravir (FVP) is an antiviral drug that was initially developed for influenza by Toyama Chemical. It selectively inhibits the RNA polymerase of RNA viruses, thus preventing viral reproduction. It displays antiviral activity against alpha-, filo-, bunya-, arena-, flavi-, and noroviruses [6, 7], as well as being active against the influenza virus.
After a pilot trial by Zhongnan Hospital of Wuhan University has found a better recovery rate in COVID-19 patients in the favipiravir group compared to the arbidol group [8], FVP is considered to be worth further investigation as a potential candidate drug for this disease.
According to the literature search, there are two published high performance liquid chromatography (HPLC) methods for determining FVP assay and impurities in active pharmaceutical ingredients [9, 10]. In both of these methods, a gradient HPLC mode was used for chromatographic separation and the run time was 60 min. FVP is not officially available in any pharmacopoeia and there is still a need for validated HPLC methods to determine FVP in pharmaceutical formulations.
REQUIRED CHEMICAL FOR TEST
Analytical grade chemicals were used without further purification in this study. Potassium dihydrogen phosphate (99.5–100.5%, Sigma–Aldrich), ortho-phosphoric acid (≥85%, Sigma–Aldrich), and HPLC-grade acetonitrile (≥99.9%, Sigma–Aldrich) were used. Deionized water was purified by a Milli-Q system (Millipore) with conductivity lower than 18.2 μS cm−1. FVP bulk powder and tablets (favicovir, 200 mg) were obtained from Atabay Pharmaceuticals and Fine Chemicals Inc (Istanbul, Turkey).
Stock standard solution
One hundred milligram pure drug was accurately weighed, dissolved in about 30 mL of deionized water and transferred to a 100 mL volumetric flask. Then the volume was completed to 100 mL with deionized water to obtain 1 mg mL−1 of stock solution. The resulting stock solution was sonicated and filtered through a 0.45 µm filter. The stock solution was further diluted with deionized water to obtain the required concentration of standard solutions (10–100 μg mL−1) before being injected into the system for analysis.
Sample solution
Ten FVP tablets were accurately weighed and transferred to a dry and clean mortar, then ground into a fine powder. Next, tablet powder equal to 250 mg FVP was transferred to a volumetric flask of 250 mL. About 100 mL deionized water was added and this flask was attached to a rotary shaker for 10 min. to completely disperse the ingredients. The mixture was sonicated for 30 min, diluted to volume with deionized water to give a solution containing 1,000 μg mL−1and then filtered through a 0.45 µm filter.
Determination of λmax
Standard solution (40 μg mL−1) was subjected to scanning between 200 and 800 nm on a ultraviolet (UV) spectrophotometer (Shimadzu UV-1800 spectrophotometer). λmax was obtained from the UV spectrum of standard solution.
Chromatographic conditions
Chromatographic analysis was performed on a column of Inertsil ODS-3V C18 (4.6 mm × 250 mm, 5.0 μm). The mobile phase consisted of potassium dihydrogen phosphate 50 mM (pH 2.3) and acetonitrile (90:10, v/v). The mobile phase was filtered and degassed through a 0.45 μm membrane filter before use and then pumped at a flow rate of 1 mL min−1. The column has been thermostated at 30 °C. The run time was 15 min under these conditions.
Method validation
The analytical method validation has been performed as per ICH guidelines of Validation of Analytical Procedure: Q2 (R1) [11, 12]. The validation parameters such as system suitability, linearity, the limit of detection (LOD), the limit of quantification (LOQ), accuracy, specificity, precision, and robustness were addressed.
Linearity
Standard calibration has been prepared using six standard solutions within the concentration range of 10–100 μg mL−1. In optimized chromatographic conditions, each standard solution was chromatographed for 15 min three times. Least squares linear regression analysis of the average peak area versus concentration data were used to evaluate the linearity of the method.
Specificity/selectivity
Selectivity is the ability of the analytical method to produce a response for the analyte in the presence of other interference. The selectivity of the method was tested by comparing the chromatograms obtained for FVP standard, tablet, and blank solutions. The parameters retention time and tailing factor were calculated in order to prove that the method chosen was specific.
Limit of detection and limit of quantification
These values were determined using the standard error (s) and slope of the regression line (m) as shown in following equations:
Precision
Precision was analyzed by calculating variations of the method in intraday (repeatability performed by analyzing standard solution on the same day) and inter-day (repeatability carried out by analyzing standard solution on three different days). Precision study was performed by injecting six times of standard solution at three different concentrations, 20, 40, and 60 μg mL−1 on the same day and three consecutive days.
Accuracy
Recovery studies were conducted by the standard addition technique to confirm the accuracy of the proposed method. In this method, 80, 100 and 120% of three different levels of pure drug were added to the previously analyzed sample solutions, and favipiravir recovery was calculated for each concentration.
Robustness
A robustness analysis was performed to determine the impact of minor yet systematic differences in chromatographic conditions. The modifications include different flow rates of the mobile phase (±0.1 mL min−1), acetonitrile ratio in the mobile phase (±1%) and column temperatures (±2 °C). After each change, System suitability parameters were checked by injecting the sample solution into the chromatographic system and the results were compared with those under the original chromatographic conditions.
Analysis of marketed formulations
Four milliliter of above prepared sample solution has been transferred into a volumetric flask of 100 mL and filled the mark with deionized water to prepare at the concentration of 40 μg mL−1 sample solution. This sample solution was filtered using 0.45 μm filter and then analyzed.
Solution stability
The stability of sample and standard solutions was monitored over a 24 h period. For this, standard and sample solutions were injected into the system at 8 h periods, and the peak area and retention time were evaluated. During the stability study, standard solutions have been stored at ambient temperature (25 °C) and protected from light.
Result and discussion
Determination of λmax
The wavelength corresponding to maximum absorbance (λmax) was determined as 323 nm from the UV spectrum of standard solution (Fig. 2).
Method development
Several preliminary studies were conducted to optimize the chromatographic conditions for the quantification of FVP. Mobile phases consisting of several buffer systems were tried at the beginning of the study; they could not meet the required system parameters. Then only potassium dihydrogen phosphate buffer system was tested without using organic modifiers, long analysis times were obtained. Different acetonitrile solution ratios were investigated to obtain optimum conditions. The acetonitrile ratio was determined as 10% against 50 mM potassium dihydrogen phosphate solution (pH = 2.3) due to the favipiravir peak being well shaped and symmetrical using this system. Eventually, it was found that the mobile phase consisting of 50 mM potassium dihydrogen phosphate (pH: 2.3 with ortho-phosphoric acid) and acetonitrile (90:10, v/v) provided stronger theoretical plates (>2,000) and peak tailing factor (<1.0). Mobile phase running at different flow rates (0.5–1.5 mL min−1) and containing mixtures of organic solvents and phosphate buffers, with ionic strengths and pH ranges were tested. Collectively, the best chromatographic conditions were achieved using an isocratic mobile phase comprising 50 mM potassium dihydrogen phosphate (pH = 2.3)-acetonitrile (90/10, v/v) at a flow rate of 1.0 mL min−1 on an Inertsil ODS-3V C18 column (4.6 mm × 250 mm, 5.0 μm) that was kept at 30 °C. The analysis was conducted at 30 °C, which offers a lot of advantages such as good chromatographic peak shape, enhanced column efficiency, and low-column pressure, in addition to being economic. The eluate was monitored using a UV detector set at 323 nm. Under the chromatographic conditions FVP were eluted at retention times 7.696. The tablet solution was analyzed for 60 min to ensure that there were no matrix components remaining in the column for much longer under the specified conditions. However, continuing the analysis after 15 min will increase both the analysis time and the cost. Overlapping peaks were not observed to overlap in samples from sample analyses injected into the system consecutively with 15 min of analysis time. Due to all these, the analysis time was determined as 15 min.
Method validation
Linearity
The stock standard solution of FVP was diluted appropriately with deionized water to obtain standard solutions within the concentration range of 10–100 μg mL−1. Each standard solution was injected three times into the HPLC system under the above-mentioned chromatographic working conditions. Linearity of the proposed method has been estimated at 6 concentration levels in the range of 10–100 μg mL−1 by regression analysis. The calibration curve was developed by plotting average peak area versus standard concentration (Fig. 3). The correlation coefficient, slope, and intercept of the regression line were determined using the least squares method. The relation between mean peak area Y (n = 3) and concentration, X expressed by equation Y = a + bX, was linear. Values of slope, intercept, and correlation coefficient (r) were 49.122, 82.598 and 0.9999, respectively as shown in Table 1. Lack of fit test was performed to evaluate linearity. In the Lack of fit test, P value (0.0516) greater than 0.05 indicates that the data satisfies the linearity condition. Overlay chromatogram of FVP standard solutions (10–100 μg mL−1) was demonstrated in Fig. 4A.
Parameter | Value |
---|---|
Linearity range (μg mL−1) | 10–100 |
Slope | 49.122 |
Intercept | 82.598 |
Correlation coefficient | 0.9999 |
Lack of fit | |
F | 2.90 |
P | 0.0516 |
SE of intercept | 6.9024 |
SD of intercept | 15.4340 |
LOD/LOQ (μg mL−1) | 1.20/3.60 |
A) Overlay chromatogram (standard
solutions, 10–100 μg mL−1, λ: 323 nm). (B) Chromatogram (standard
solution, 80 μg mL−1, λ: 323 nm). (C) Chromatogram (sample solution,
40 μg mL−1, λ: 323 nm). (D) Chromatogram (Blank solution, λ:
323 nm)
Specificity/selectivity
The chromatogram of FVP standard
solution has been given in Fig. 4B. There is only one peak at the
retention time of 7.696 min. The chromatogram of the tablet solution has been
given in Fig. 4C. There is only one peak at the
retention time of 7.696 min in this chromatograme. There are no other peaks
caused by excipients and additives in this chromatograme. The chromatogram of
the mobile phase has also given in Fig. 4D. There are no other peaks caused by
contents of the mobile phase in this chromatograme. This indicates that the
analytical method is specific. The parameters retention time and tailing factor
were calculated in order to prove that the method chosen was specific.
Retention time, theoretical plate number, and peak tailing factor values were
7.696, 13798, and 0.920, respectively. All of the values were within the
accepted level.
Precision
Precision study was performed by
injecting six times of standard solution at three different concentrations, 20,
40, and 60 μg mL−1 on the same day and three consecutive days. The
precision data were given in Table 2. All RSD values for retention time and
peak area for selected FVP concentrations were less than 0.5 and 2.0%,
respectively. In this case, the method is precise and can be used for our
intended purpose.
Std. conc. |
Intraday precision |
Interday precision |
||||
Found conc. (6) |
Peak area RSD (%) |
Retention time RSD
(%) |
Found conc. (6) |
Peak area RSD (%) |
Retention time RSD
(%) |
|
20 |
20.02 |
0.178 |
0.015 |
19.92 |
0.195 |
0.034 |
40 |
40.35 |
0.066 |
0.004 |
40.08 |
0.082 |
0.012 |
60 |
60.24 |
0.041 |
0.002 |
60.10 |
0.064 |
0.025 |
Accuracy study
A known quantity of standard
solution has been added to the sample solutions previously analyzed at three
different levels (80%, 100% and 120%). The amount recovered for favipiravir has
been calculated for three concentration. The recovery data were summarized
in Table 3. Percent RSD values for all analyses
were less than 2% indicating that excipients found in pharmaceutical
formulations do not interfere and analytical method is very accurate.
Table 3.
Recovery data
Spiked
level (%) |
Amount
added (μg mL−1) |
Amount
recovered (μg mL−1) |
Recovery
(%) |
Average
(%) |
SD |
RSD
(%) |
80 |
32 |
31.78 |
99.31 |
99.96 |
0.157 |
0.157 |
32 |
31.74 |
99.19 |
||||
32 |
31.84 |
99.50 |
||||
100 |
40 |
39.79 |
99.48 |
99.93 |
0.180 |
0.180 |
40 |
39.75 |
99.38 |
||||
40 |
39.89 |
99.73 |
||||
120 |
48 |
48.02 |
100.04 |
99.75 |
0.125 |
0.125 |
48 |
48.08 |
100.17 |
||||
48 |
47.96 |
99.92 |
Robustness
The results showed that the change
in flow rate and mobile phase concentration had little effect on the
chromatographic behavior of FVP. The small change in the mobile phase flow rate
and acetonitrile content have a small impact on the retention time of FVP. The
change in the column temperature did not have a significant effect on the
method. The results of this study, expressed as % RSD, were presented in Table 4.
Condition |
Variation |
Assay |
SD |
RSD |
|
Mobile phase flow
rate (1.00 mL min−1) |
0.90 mL min−1 |
99.86 |
0.60 |
0.60 |
|
1.10 mL min−1 |
99.94 |
0.62 |
0.62 |
||
Acetonitrile ratio
in mobile phase (10%) |
9% |
100.12 |
0.67 |
0.67 |
|
11% |
99.96 |
0.71 |
0.71 |
||
Column temperature
(30 °C) |
28 °C |
99.96 |
0.34 |
0.34 |
|
32 °C |
100.05 |
0.32 |
0.32 |
Solution stability
The stability of sample and standard
solutions was monitored over a 24 h period. For this, standard and sample
solutions were injected into the system at 8 h periods, and the peak area and
retention time were evaluated. No changes in standard concentrations have been
observed over a period of 24 h. The % RSD for peak area (n = 3) was
0.275% and the value for retention time (n = 3) was 0.12% for
standard solution. The results have been demonstrated in Table 5. No major changes in active ingredient
concentration have also been found in the tablet solution.
Table 5.
Standard solution stability (40 μg mL−1)
Time (h) |
Peak area |
Mean |
SD |
RSD (%) |
Retention time (min) |
Mean |
SD |
RSD (%) |
8 |
2013.5 |
2019.7 |
5.5 |
0.275 |
7.594 |
7.596 |
0.009 |
0.115 |
2021.3 |
7.606 |
|||||||
2024.2 |
7.589 |
|||||||
16 |
2025.6 |
2020.0 |
5.2 |
0.257 |
7.597 |
7.599 |
0.015 |
0.192 |
2019.2 |
7.614 |
|||||||
2015.3 |
7.585 |
|||||||
24 |
2012.2 |
2018.1 |
6.5 |
0.323 |
7.610 |
7.597 |
0.011 |
0.150 |
2025.1 |
7.588 |
|||||||
2017.1 |
7.594 |
Application of the
method to the marketed tablets
The developed and validated method
has been applied successfully for determination of FVP in pharmaceutical
formulations. The result of assay of the marketed tablet of favipiravir is
shown in Table 6. The results obtained are closely
related to the amount indicated on the labels of the tablets. This shows that
the method for content evaluation is useful.
Method application results
Formulation | Label claim (mg) | Amount of drug (mg) | % Assay ± SD |
Favicovir tablet | 200 | 200.35 | 100.18 ± 0.38 |
Conclusions
A very quick, cost-effective, precise and accurate HPLC method for the determination of FVP has been developed and validated in compliance with ICH guidance Q2. Besides the short run time (15 min), retention time (7.696) and flow rate of mobile phase (1 mL min−1) made the method attractive because these features save analysis time and cost. Potassium dihydrogen phosphate, used as a general purpose buffer, has many interesting properties. The most important of these features are good buffering capacity in the selected pH range, easy availability, low toxicity and cost, and greatly improved separation ability without colon degradation. In short, this method is sensitive, selective, reproducible and rapid for favipiravir in bulk and tablets. The accuracy and precision are within reasonable limits, the maximum of quantification is as small as 3.60 μg mL−1 and finally analytical method is reliable and robust.