Buffer and buffer solution , work of buffer in mobile phase complete Guidlines , Why is pH Important?

 What is Buffer and Buffer solution 

Buffer  A solution that resists a change in pH when small amounts of acid or alkali are added to it, or when it is diluted with water.

Buffer solution  An aqueous solution consisting of a mixture of a weak acid and its conjugate base or a weak base and it's conjugate acid [i.e., CH3COOH acetic acid and CH3COO acetate ion from sodium acetate (CH3COONa)]

In a buffer solution there is an equilibrium between a weak acid, HA, and its conjugate base, A (or vice versa as stated above).

HA + H2OH3O+ + A

 

 When an acid (H+ or H3O+ ) is added to the solution, the equilibrium moves to the left, as there are hydrogen ions 

(H+ or H3O+ ) on the right-hand side of the equilibrium expression.

When a base [hydroxide ions (OH )] are added to the solution, equilibrium moves to the right, as hydrogen ions 
(H+ or H3O+ ) are removed in the reaction

H+ + OHH2O

 

Thus, in both cases, some of the added reagent is consumed in shifting the equilibrium in accordance with Le Chatelier’s principle, and the pH changes by less than it would if the solution were not buffered.


Why Do We Need a Buffer Solution ?

The mobile phase pH can change on standing, with ingress of CO2 from the atmosphere for example, and a buffer can help to combat this effect to a certain extent. Similarly, volatile reagents, such as TFA, may also selectively evaporate, thus changing the eluent pH. There is, however, no substitute to regularly replacing the buffer on our HPLC system!

Perhaps the largest potential for pH change is on mixing of the injection slug with the mobile phase within the tubing and components of the autosampler, or at the head of the HPLC column, where more extensive mixing of the sample diluent and eluent occurs. If the sample diluent pH differs greatly from the eluent, the ‘local’ pH will change as the two mix — leading to retention time variability and peak distortion as not all analyte molecules experience the same solution pH and, therefore, may exhibit different partitioning behaviour. 


 

Why is pH Important?
The pH of the mobile phase effects the retention time of ionizable analytes.
Altering the mobile phase pH may alter the extent to which the analytes are ionized, affecting their relative hydrophobicity, the extent to which they interact with the stationary phase and hence their retention time. Ultimately, changes in pH will lead to changes in retention time for ionizable compounds, which may in turn lead to selectivity changes within the chromatogram.

Choosing the Right Buffer/Buffer Capacity
Many factors influence the choice of buffer, but the two major considerations tend to be:

  • What is the required pH of the mobile phase (which is dictated by the analyte properties).
  • Does the buffer need to be volatile (which usually depends on whether one is using mass spectrometric detection).

The mobile phase pH value will depend upon the analytes pKa (partial acid dissociation constant) value and may be derived by experimentation or computer simulation or both. For a weak acid in solution, we can write the equilibrium,

HAA− + H+ 

to represent the dissociation of the acid.

The dissociation constant (Ka) for this equilibrium can be written as;

 

Because of the wide range of Ka values possible (several orders of magnitude), it is more convenient to talk about the partial acid dissociation constant, which can be written
as;

pKa = −log10Ka

We can also express the pKa in terms of pH using the Henderson Hasselbalch equation;

It should be obvious that pH is equal to pKa when the associated and dissociated forms of the acid are present in equal concentrations (i.e., when [A ] = [HA]). This represents the value at which the acid (or base) is 50% ionized [dissociated if acidic (A ), associated if basic (BH+ )].

A buffer is chosen, so that the buffer pKa is as close to the required eluent pH as possible, and certainly within one pH unit of this value (see Table 1). When this condition is satisfied, the buffering capacity of the solution is at its maximum and a more robust method will result, using a lower concentration of the buffer. When the eluent is ±1 pH unit from the buffer pKa, the buffering capacity has already fallen to 33% of the capacity obtained when the eluent pH = buffer pKa. 

If the buffer is incorrectly chosen, it will need to be added at much higher concentrations to be effective, which will lead to method robustness issues and changes in the selectivity of the separation that are more difficult to predict and manipulate.

So, from Table 1, if an eluent pH of 4.2 is required to achieve a particular separation, the ideal buffer system to chose would be ammonium acetate (pKa 4.76) adjusted to pH 4.2 using acetic acid or ammonium formate (pKa 3.74) adjusted to pH 4.2 using formic acid.

It should be noted that although the phosphate species have three suitable pKa values, NONE are suitable for use within the range pH 3 to 6!



No comments:

Post a Comment

COVISHIELD™ Overview , You should know.....

  ChAdOx1 nCoV- 19 Corona Virus Vaccine (Recombinant) COVISHIELD™                                                                           ...

Most Popular Post